MODEL | HVS100-2x60 | HVS100-120 | HVS100-5 | HVS100-40 | HVS100-80 | HVS100-150 | HVS100-4x60 | HVS100-2x120 |
Maximal high voltage output options | +/-1.4 kV +/-2.6 kV |
1.8 kV 2.6 kV 3.1 kV 3.6 kV |
1.8 kV |
1.3 kV 1.8 kV 2.5 kV 3.6 kV 4.0 kV |
1.8 kV 2.6 kV 3.1 kV 3.6 kV 4.0 kV |
1.8 kV |
+/-1.4 kV +/-2.0 kV +/-2.6 kV +/-3.6 kV |
1.8 kV |
Maximal output power at maximal output voltage | 2 x 60 W | 120 W | 5 W | 40 W | 80 W | 150 W | 4 x 60 W | 2 x 120 W |
Voltage control limits | 40% Umax to Umax | |||||||
Remaining power at the output of 24 V DC for external needs | 55 W | 55 W | 190 W | 150 W | 100 W | Not available | ||
Mains voltage | 90 to 264 V AC, 47 - 63 Hz | 24 V DC1) | ||||||
Dimensions (W x L x H) | 230 x 245 x 53.5 mm | |||||||
Weight | 2.5 kg |
1) External desktop AC-DC adapter unit: weight 1.6 kg, dimensions 112 x 222 x 45 mm (included).
Maximal power is proportionally lower by tuning to lower HV output.
Output ripple is 0.5% typically at maximal power for all models of HV power supplies.
Laboratory type HVS100 series high voltage power supply is especially designed for powering of drivers of Pockels cells. However, it can also be used as a general-purpose high voltage power supply for other electronic devices and different applications in the laboratory.
HVS100 features two voltage output channels – adjustable high voltage output channel and 24 V DC output channel. Required high voltage value can be set and adjusted in the limits from 40% of maximal specified voltage to maximal specified voltage (40% Umax to Umax).
HVS100 has 2.4-inch OLED display for high voltage control and current consumption monitoring at the output channel.